Operating Models |
MPP1000 stand-alone unit Single-probe/detector version or dual-probe/detector version |
range |
0 to 99.9ms |
Resolution |
0.1ms |
Inputs |
2(left/light) for acoustic pickups |
Outputs |
1 jack for headphone, 300Ω per side |
Volume |
Headphone volume adjustable |
Acoustic Level |
Manual |
Electromagnetic Gain |
Automatic |
Acoustic Bands |
125 to 1000Hz |
Acoustic Pickup |
6-ft(1.82m)cord |
Display |
LCD with switchable backlight |
Power |
8 standard AA cell batteries Automatic shutdown after 1 hour |
Battery Life |
±65hrs. continuous usage, alka-line; ±85hrs continuous usage, lithium (Equates to several weeks/months of normal usage) >200 hours, intermitten, less with backlight enabled |
Operating temperature range |
-4 to 122oF(-20 to +50 oC) |
Storage temperature range |
-40 to 158oF(-40 to +70 oC) |
Invironmental |
Rated to IP54 |
Humidity |
<95% noncondensing |
MPP1000 Dimensions |
8 L x 6.5 W x 3.25 H in 203 L x 165 W x 83 H mm |
MPP1000 Weight |
2.15lb(98kg) |